- For all DELTA 30 type fume hoods
- Exhaust air volume is controlled according to the free inflow area, both with vertical, horizontal or combined front sash opening as well as side sash opening
- Energy savings and increased safety at work without compromising ease of use
- System solutions for achieving optimal room air balance
- Controlled electrically or pneumatically
- Fast electronic control with full integration into building process control technique, via LON bus technology or conventionally via analogue signals
- Direction sensitive inflow sensor ensures constant laminar flow conditions and automatic discharge of thermal loads in the interior of the fume hood
- Vmax switch for increasing exhaust air volume in the event of an emergency, particularly with the front sash closed
- Real system solution, in which all exhaust air, inlet air, drive, and control components are matched to each other
- Automatically closing front sash for increased protection and safe realisation of the planned room air balance, with movement detector and options such as foot-operated switch and light barrier available as optional extra
|
|

 |